Non-Standard Low Temperature Low Pressure Test Chamber

2 years ago
107
Share
Non-Standard Low Temperature Low Pressure Test Chamber.

Product: Non-standard low temperature low pressure test chamber
Inner Size (mm) : 450×450×500mm (W×D×H)
Water Spray : can control the spray time from 0.1s to 10s
Steam Injection: can control the injection time from 0.1s to 10s
Limit Temperature Range: -65℃~Room temperature
Air pressure Range: Normal pressure~10Pa continuously adjustable
Depressurization Rate: Normal pressure~10Pa≤40min
Cooling Rate: Room Temperature~-60℃≤60min

Non-standard windows are symmetrically on the left and right sides of the chamber
with high-transmittance quartz glass (light transmission> 90%, and the thickness is about 20mm )
the surface is coated with anti-fog treatment.

The inner chamber is sprayed with black aerospace black paint.

Water mist and steam nozzles are arranged at the bottom of the chamber unit and inserted into the chamber through an trepanning.

The connection part of the spray pipe and the chamber unit is welded and sealed, and the spray pipe inside the chamber is treated with an electric heating belt and heat preservation method to avoid low temperature ice blocking .
The diameter of the spray pipe inside the chamber is 4-6mm stainless steel pipe, the thickness is not less than 0.6mm, and the pressure resistance is above 6Mpa.
The spray direction can be fine-tuned by bending the stainless steel pipe.